What is the difference between CMM calibration and verification?

Calibration states the relationship between a machine's readings and a reference value with its uncertainty. Verification confirms the machine still meets a stated specification, such as ISO 10360-2. Most work ordered as calibration is a performance verification.

Do you offer CMM calibration near me?

On-site service covers Oklahoma City and the metros listed on this site, generally within a day's drive. Machines outside that radius are typically quoted for ship-in service instead, with return freight and scheduling confirmed in the quotation.

What happens if the machine fails as-found?

As-found deviations are recorded and reported before any adjustment is made. That record supports a reverse-traceability impact assessment back to the last passing calibration.

What does ISO 10360-2 actually test?

Length measurement error, designated E0 and E150, measured at five calibrated lengths, three repeats, seven positions: 105 measurements. Probing error, designated PFTU under ISO 10360-5:2020, is tested separately against 25 points on a calibrated test sphere.

Can calibration be done on site, and how long is the machine down?

Bridge, gantry, cantilever and horizontal-arm machines are commonly tested on site. Time on site varies with machine size and axis travel; the written quotation confirms the figure for a given job.

What temperature does the room need to be?

ISO 1 sets 20 degrees C as the dimensional reference temperature. The machine manufacturer's specified environment applies to an acceptance test; the user sets the environment for a reverification, so the recorded conditions on the certificate matter.

Is the certificate accredited, and how do I check the scope?

Certificates are issued under the performing laboratory's scope of accreditation to ISO/IEC 17025:2017. The scope can be checked against the accrediting body's public directory before the work is accepted.

What is a CMC and why does it matter to my tolerances?

Calibration and measurement capability is the smallest uncertainty a laboratory can claim within its scope. Reporting an uncertainty smaller than the scope CMC is a finding, not a selling point. The CMC still needs to support a 4:1 test uncertainty ratio against the tolerance being inspected.

Will this satisfy my IATF or AS9100 auditor?

Both standards require evidence that an external laboratory is accredited to ISO/IEC 17025:2017 or otherwise acceptable to the customer, and that out-of-tolerance findings trigger an impact assessment. Certificate content and scope documentation are provided to support that review.

What is an error map, and can it be updated on my controller?

The volumetric compensation map corrects systematic geometric error across the measuring volume. It can be updated where the controller permits access; some OEM controllers lock this function to factory service only.