Primary Service
CMM Calibration and Repair
Request a quotation for performance verification or repair.
Request a quotation

What a Performance Verification Covers
Calibration establishes a machine's relationship to reference values with a stated uncertainty. Performance verification confirms the machine still meets a stated specification, usually ISO 10360-2 or ASME B89.4.1, and the trade commonly calls both calibration.
Scale inspection and cleaning on X, Y and Z come first, followed by a reader head check and confirmation of air bearing and airline condition, before any length measurement is taken. Multisensor machines, tested under ISO 10360-9:2013, follow the same sequence with an added check against each sensor type qualified on the system.
Sequence of a Visit
Geometry is checked before volumetric performance, so a squareness or straightness error is found and corrected before it contaminates the length test.
- 01
Geometry check
Straightness, linearity and squareness in the XY, ZX and ZY planes.
- 02
Length measurement error
Five calibrated lengths, three repeats, seven positions: 105 measurements under ISO 10360-2.
- 03
Probing error
25 points on a separate calibrated test sphere, per ISO 10360-5, distinct from the stylus qualification sphere.
- 04
As-left verification
Results confirmed and certificate issued after any adjustment is complete.
From Scales to Certificate
- 01
Inspect
Scales, reader heads and air bearings checked before any test runs.
- 02
Test
Length measurement error and probing error measured against calibrated artifacts.
- 03
Adjust
Geometric correction and volumetric compensation updated where the controller allows it.
- 04
Certify
As-found and as-left results issued with uncertainty and decision rule stated.

Why Oklahoma City CMM Calibration and Repair
- 01
Artifacts, not estimates
Gauge blocks to ISO 3650, step gauges and ball bars supply the calibrated lengths.
- 02
OEM-locked maps respected
Volumetric compensation is updated only where controller access permits it.
- 03
Repair scoped separately
Probe heads, drives, counterbalance and crash recovery are quoted apart from verification.
Coverage across Oklahoma and the Southern Plains
Service area covers Oklahoma City and the surrounding region. Machines and gages outside the stated radius may still be quoted; travel terms are confirmed before scheduling.
Other services
Discuss CMM Calibration and Repair Requirements
Send machine make, model and travel range for a visit quotation.
